Floor-standing models with capacities ranging from 150 to 1,200 liters

Vertical door autoclaves are powerful, compact, and cost-effective floor-standing units. Thanks to the rectangular chamber design, a large loading capacity is achieved within relatively small external dimensions. This is the optimal solution for maximizing space efficiency in your laboratory.
The autoclave can be individually configured to meet your specific sterilization requirements through a variety of available options. Both single-door versions and pass-through models featuring a gas-tight barrier are available. We look forward to working together with you to develop the perfect solution for your sterilization process.

Description
Steam Supply Options:
  • House steam: Connection to the facility's centralized clean steam pipeline.
  • Intrinsic steam generator: A built-in steam generator integrated directly into the autoclave housing.
Design Configurations:
  • Pass-through: Double-door configuration with a gas-tight barrier to separate 'dirty' and 'clean' zones (for BSL-2/BSL-3 laboratories).
  • GMP-compliant: Built in compliance with DIN 58950 and FDA standards (critical for pharmaceutical manufacturing).
Rapid Cooling Systems (for Liquids):
  • Double jacket: Cooling through the chamber walls. Combined with sterile air counterpressure, this is the fastest method to cool liquids while preventing them from boiling over.
  • Ring cooling system: Connection to the facility's circulating water system via a heat exchanger to maximize water savings.
  • Circulating air fan: An internal fan for mixing steam and air, which accelerates cooling and enables operation in steam-air mixture mode.
Special Modes for Sealed Containers:
  • HWBV (Hot Water Baiting/Vessel - Hot Water Overheated Sterilization): The fastest and most gentle method for sterilizing liquids in hermetically sealed containers.
  • Spray cooling: Spray/atomized cooling for a sharp reduction in temperature post-cycle.
Safety and Vacuum:
  • Exhaust air filter: An exhaust air filter with condensate treatment for handling infectious materials (Biosafety Levels S2/S3). Particle retention rate > 99.5% (0.02 µm).
  • Vacuum pump set: A vacuum pump for mechanical air removal (fractionated vacuum) and final drying of sterilized items.









Ergonomics and Accessories:
  • Transport and batch trolley: Loading trolleys for convenient feeding of heavy batches into the chamber.
  • Insert baskets/containers: Baskets and containers with a solid bottom and lid (designed specifically for waste disposal or load optimization).
Documentation Systems:
  • Digital batch recording and user management (Siemens): Network-compatible and fully compliant with the FDA 21 CFR Part 11 standard. This is a solution for large enterprises and pharmaceutical manufacturing, where electronic signatures and data tampering protection are critical.
  • USB flash drive: The simplest method. It allows saving up to 1,000 cycles to a flash drive, which can then be opened on any computer for verification or archiving.
  • 'Sterilog' software: Software for real-time, direct transmission of sterilization process data to a PC via an Ethernet port. Convenient for instant monitoring and generating digital reports.
Pharmaceutical qualification DQ, IQ, and OQ in accordance with the DIN 58950 standard; FDS, HDS, and SDS specifications.