CO2 incubator ICO50

The control system of the Memmert ICO CO2 incubator is so finely tuned that the appliance reaches the setpoint temperature very quickly without any temperature overshoots.
Alarm notifications sent directly to your mobile phone guarantee the safety of the chamber load. Perfect hygiene is also a matter of course: rounded corners ensure easy and thorough cleaning, and the interior can be sterilised in 60 minutes at 180 °C.

Control technology
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese.
Function SetpointWAIT: The process time does not start until the set temperature is reached.
Adjustable parameters: Temperature (Celsius or Fahrenheit), CO2 concentration, programme time, time zones, summertime/wintertime.
Interfaces and communication
Interface: Ethernet LAN, USB.
Documentation: Programme stored in case of power failure.
Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port.
Safety control system
AutoSAFETY system: Additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value within a preset tolerance range; alarm in case of over- or undertemperature, heating function switched off in case of overtemperature, cooling function switched off in case of undertemperature.
Autodiagnostic system: Integrated fault diagnosis for temperature and CO2 control.
Alarm: Visual and acoustic.
Heating concept
6-sided heating system: Multifunctional large-area heating on four sides of the chamber, supplemented by independent door and rear wall heating to prevent condensation.
Standard equipment
Calibration certificate (with humidity): Works calibration certificate for +37 °C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for appliances with active humidity control.
Calibration certificate (with oxygen): Works calibration certificate for +37 °C, 5 % CO2, 90 % rh and 10 % O2 (requires options K7 and T6); standard equipment for appliances with O2 control.
Calibration certificate (standard): Works calibration certificate (in the geometric centre of the chamber) at +37 °C, 5 % CO2 for standard models.
Door (outer): Fully insulated stainless steel door with 2-point locking (compression door lock).
Door (inner): Glass inner door with opening (Ø 8 mm) to take gas samples.
Gas connection set: Basic connection set for CO2 (hose with connection coupling and clamp).
Standard accessories: Membrane filter (all incoming gases pass through it before reaching the chamber to remove impurities and pollutants).
Internals: 1 perforated stainless steel shelf.
Dish: 1 stainless steel water dish.
Dimensions and chamber construction
Interior material: Corrosion-resistant stainless steel 1.4301 (ASTM 304).
Volume: 56 l.
Internal dimensions (W x H x D): 400 x 425 x 330 mm (depth less 35 mm for fan).
Max. number of internals: 5 pcs.
Max. total loading of chamber: 75 kg.
Max. loading per internal: 15 kg.
Housing: Textured stainless steel.
External dimensions (W x H x D): 559 x 795 x 521 mm (depth + 56 mm for door handle).
Rear panel of housing: Zinc-plated steel.
Electrical data
Option 1: Voltage 230 V, 50/60 Hz | Electrical load approx. 1100 W.
Option 2: Voltage 115 V, 50/60 Hz | Electrical load approx. 1100 W.
Installation and ambient conditions
Installation requirements: The distance between the wall and the rear of the appliance must be at least 15 cm; the clearance from the ceiling must not be less than 20 cm; the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature: +10 °C to +35 °C.
Humidity: Relative humidity max. 70 %, non-condensing.
Installation altitude: Max. 2,000 m above sea level.
Overvoltage category: II.
Pollution degree: 2.
Packing and shipping data
Transport information: The appliance must be transported upright.
Customs tariff number: 8419 8998.
Country of origin: Federal Republic of Germany.
WEEE-Reg.-No.: DE 66812464.
Packaging dimensions (W x H x D): approx. 730 x 950 x 640 mm.
Net weight: approx. 55 kg.
  • Gross weight (in carton): approx. 74 kg.
TECHNICAL SPECIFICATIONS
Temperature
Working temperature range: 5 °C above ambient temperature up to +50 °C.
Standard sterilisation programme: 60 minutes at 180 °C (without removing the sensors).
Setting temperature range: +18 to +50 °C.
Setting accuracy temperature: 0.1 °C.
Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring and taking over functions in case of an error.
Spatial temperature distribution in chamber: at +37 °C: +/- 0.3 K.
Temperature fluctuation over time: at +37 °C: +/- 0.1 K.
Humidity
Humidity control (standard): Humidity limitation by Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %.
Setting accuracy humidity: 0.5 % rh.
Setting range active humidity control (with option K7): 40 to 97 % rh or rh-Off.
Control of standard atmosphere components
CO₂ control: Digital electronic CO₂ control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation.
Setting range CO₂: 0 to 20 % CO₂.
Setting accuracy CO₂: 0.1 %.
Variation in time CO₂: +/- 0.2 % CO₂.
Setting range O₂ (with option T6): 1 to 20 % O₂.
Setting accuracy O₂: 0.1 % O₂.

Product specification