Parameter; Value
Built-in ionizer; Yes
Maximum capacity [Max]; 610 g
Readability [d]; 0.01 mg
Calibration; Automatic internal
Pan size; Ø 80 mm
Linearity; ± 0.0005 g
Stabilization time (sec.); ≤ 4
Repeatability; 0.1 mg (≤ 220 g) / 0.3 mg (220–310 g) / 0.35 mg (310–610 g)
Verification; N/A
Minimum load; 10 mg
Display; LCD
Tare range (subtractive); 610 g
Span drift (+10...+30°C); ± 2 ppm/°C
Interface; RS232
Draft shield type and dimensions; 3 sliding doors / 162 x 171 x 225 mm
Operating temperature; 5 – 35°C (15–30°C for verified models)
Power supply; 110-230 V AC, 50/60 Hz; output 24 V, 500 mA, 13 VA
Power consumption (max.); 2 W
Calibration weight (class); 200 g (E1)
Balance dimensions (W x D x H); 215 x 345 x 345 mm
Net / gross weight (approx.); 7.2 / 9.8 kg
Packaging dimensions (W x D x H); 390 x 520 x 630 mm