Constant climate chamber HPP2200eco

Maximum energy efficiency in continuous operation – tailor-made for stability testing in accordance with ICH guidelines, stability tests for cosmetics and food as well as environmental tests and material tests.

The HPPeco climate chamber with Peltier cooling is designed for stability tests according to ICH (Q1A) standards, as well as testing of cosmetics and food products. Thanks to the operating range from the freezing point, the chamber is suitable for material testing and environmental tests in industry, while the use of the P1 option is recommended. For detailed information about the P1 option, please visit the manufacturer's website.












Control technology
  • ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
  • Language setting: German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese.
  • Adjustable parameters: Temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime.
  • Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days.
  • Function HeatBALANCE: Adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %.
  • Function SetpointWAIT: The process time does not start until the set temperature is reached.
  • Calibration: Three freely selectable values each, temperature and humidity.
  • Communication: Interface Ethernet LAN, USB.
  • Documentation: Programme stored in case of power failure.
  • Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port.
Safety
  • Temperature control: Over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display.
  • AutoSAFETY: Additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature.
  • Autodiagnostic system: Integral fault diagnostics for temperature and humidity control.
  • Alarm: Visual and acoustic.
Heating concept
  • Peltier: Energy-saving Peltier heating/cooling system integrated in the rear (heat pump principle).
Standard equipment
  • Standard works calibration certificate: +25 °C / 60 % rh, +40 °C / 75 % rh.
  • Door: Stainless steel doors with glass sectors, fully heated inner glass panes integrated in the full-sight glass door with 2-point locking (compression door lock).
  • Internals: 6 stainless steel grid(s), electropolished.
  • Standard accessories: Water tank including connection hose.
Stainless steel interior
  • Dimensions (W x H x D): 1972 x 1450 x 750 mm (D less 32 mm for fan - Peltier).
  • Volume: 2140 l.
  • Max. number of internals: 42.
  • Max. loading of chamber: 330 kg.
  • Max. loading per internal: 30 kg.
Textured stainless steel casing
  • Dimensions (W x H x D): 2157 x 1913 x 905 mm (D +56 mm door handle & +111 mm Peltier element).
  • Installation: On lockable castors, adjustable in height.
  • Housing rear: Zinc-plated steel.
Electrical data
  • Voltage / Electrical load: 230 V, 50/60 Hz / approx. 2,000 W.
Ambient conditions
  • Set Up: The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
  • Ambient temperature: 16 ºC to 40 ºC.
  • Humidity rh: max. 70 %, non-condensing.
  • Altitude of installation: max. 2,000 m above sea level.
  • Overvoltage category: II.
  • Pollution degree: 2.
Packing/shipping data
  • Transport information: The appliances must be transported upright.
  • Customs tariff number: 8419 8998.
  • WEEE-Reg.-No.: DE 66812464.
  • Dimensions approx incl. wooden box (W x H x D): 2300 x 2200 x 1190 mm.
  • Net weight: approx. 474 kg.
  • Gross weight wooden box: approx. 718 kg.

TECHNICAL SPECIFICATIONS
Temperature
  • Setting temperature range without light, with humidity: +15 to +60 °C
  • Setting temperature range without light, without humidity: 0 to +70 °C
  • Working temperature range without light, with humidity: +15 °C (at least 10 °C below ambient temperature) to +60 °C
  • Working temperature range without light, without humidity: 0 °C (at least 20 °C below ambient temperature) to +70 °C
  • Setting accuracy temperature: 0.1 °C
  • Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity
  • Setting range humidity: 10 – 80 % rh
  • Humidity supply: humidity supply with distilled water from external tank by self-priming pump
  • Humidification: humidification by hot steam generator
  • Dehumidification: dehumidification by cold trap using the Peltier technology
  • Setting accuracy humidity: 0.5 % rh
Product specification