Climate chamber ICH260L

Unparalleled homogeneity of temperature and humidity. For stability testing according to ICH – WHO – EMA – ASEAN – GMP – GLP – GCCP.


The humidity of the load, the ambient conditions and the respective temperature-humidity working range are decisive factors when choosing the right appliance. In the adjacent diagram, you can see the possible temperature-humidity combinations for our climate chamber ICH. Within the stated temperature and humidity range, long-term continuous operation without condensation is possible. The extent to which condensation may occur in threshold ranges depends on the moisture content of the load and the ambient conditions.











Control of standard components
  • Light programming: Daylight (white) and UV light can be programmed independently of each other via the controller.
  • Uniformity of illumination: ±6 % for cold white light, ±10 % for UV light.
  • Illumination: Light illumination unit consisting of 4 fluorescent lamps with cold white light (standard illuminant D65, 6,500 K) and 2 UV lamps in the spectral range of 320–400 nm.
Control technology
  • ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
  • Language setting: German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese.
  • Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days.
  • Function SetpointWAIT:
  • The process time does not start until the set temperature is reached.
  • Adjustable parameters: Temperature (Celsius or Fahrenheit), fan speed, illumination, programme time, time zones, summertime/wintertime.
Ventilation
  • Fan: Forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps.
Communication
  • Documentation: Programme stored in case of power failure.
  • Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port.
Safety
  • Temperature control (mechanical): Mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10 °C above nominal temperature.
  • Temperature control (electronic): Over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display.
  • AutoSAFETY: Additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature.
  • Autodiagnostic system: Integral fault diagnostics for temperature and humidity control.
  • Alarm: Visual and acoustic.
Heating concept
  • Air jacket: Air jacket heating system for gentle all-around heating.
  • Heating concept: No drying-up of the load caused by the cooling device due to separation of thermal jacket from interior.
  • Defrosting: Highly efficient automatic defrosting system.
Standard equipment
  • Works calibration certificate: Standard values +10 °C and +37 °C, 60 % rh at +30 °C.
  • Entry port: Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed by a silicone stopper, standard position at the back.
  • Internals: 2 stainless steel grid(s), electropolished.
  • Standard accessories: Water tank including connection hose.
Stainless steel interior
  • Dimensions (W x H x D): 640 x 800 x 500 mm (D less 33 mm for fan).
  • Volume: 256 l.
  • Max. number of internals: 9.
  • Max. loading of chamber: 200 kg.
  • Max. loading per internal: 20 kg.
Textured stainless steel casing
  • Dimensions (W x H x D): 824 x 1552 x 685 mm (D +56 mm door handle).
  • Installation: On lockable castors.
  • Housing rear: Zinc-plated steel.
Electrical data
  • Voltage / Electrical load: 230 V, 50/60 Hz / approx. 1,450 W.
  • Voltage / Electrical load: 115 V, 50/60 Hz / approx. 1,450 W.
Ambient conditions
  • Set Up: The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
  • Ambient temperature: 15 °C to 28 °C (up to 34 °C with limited temperature range).
  • Humidity: rh max. 70 %, non-condensing.
  • Altitude of installation: max. 2,000 m above sea level.
  • Overvoltage category: II.
  • Pollution degree: 2.
Packing/shipping data
  • Transport information: The appliances must be transported upright.
  • Customs tariff number: 8419 8998.
  • Country of origin: Federal Republic of Germany.
  • WEEE-Reg.-No.: DE 66812464.
  • Dimensions approx incl. carton (W x H x D): 930 x 1760 x 930 mm.
  • Net weight: approx. 160 kg.
  • Gross weight carton: approx. 217 kg.
TECHNICAL SPECIFICATIONS
Temperature
  • Working temperature range without humidity: 0 to +60 °C
  • Setting temperature range: 0 to +60 °C
  • Working temperature range with humidity and/or light: from +10 °C up to +60 °C
  • Setting accuracy temperature: 0.1 °C
  • Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity
  • Humidity supply: humidity supply with distilled water from external tank by self-priming pump
  • Humidification: humidification by hot steam generator
  • Humidity control: active humidifying and de-humidifying adjustable from 10 – 80 % rh with digital display of relative humidity – resolution of display 0.1 %, setting accuracy 0.5 %