Laboratory and benchtop autoclaves

Our benchtop autoclaves not only fit on any laboratory bench, but their chamber is also specially designed to achieve optimal loading capacity when using a variety of standard flasks (such as Schott) and Erlenmeyer flasks. Our laboratory and benchtop autoclaves are available in six different sizes and are engineered to save space.

Description of Laboratory and Benchtop Autoclaves
A key feature of the LabStar 40 and LabStar 70 models is the square (rectangular) chamber. It provides a much larger loading capacity compared to devices with a round chamber, despite having nearly identical external dimensions. Starting with the LabStar 100 model, these laboratory autoclaves are also available in pass-through configurations.

Description of Individual Components
  • Steam Generator
  • Integrated into the autoclave housing and operates independently of the sterilization chamber. It ensures rapid heating and cooling compared to conventional heating systems.
  • Rapid Water Cooling with Fan
  • Reduces the cooling time for liquids by more than 80% compared to natural cooling. This is achieved through an internal heat exchanger ('rapid cooler') with continuous water flow.
  • Vacuum Pump Kit
  • Designed for mechanical air removal and vacuum drying. It supports single-stage, fractionated vacuum processes, as well as continuous or pulsating vacuum drying. The process can be accelerated by heating the chamber jacket.
  • Exhaust Air Filtration
  • A condensate sterilization and filtration system for handling pathogenic substances in Biosafety Level S2/S3 (BSL-2/BSL-3) laboratories in accordance with TRB 100.
  • Steam-Air Mixture Process
  • For the sterilization of tightly sealed vials and pressure-sensitive items. An additional temperature sensor, deflectors, and a fan ensure precise control and temperature distribution.
  • Documentation
  • Options include a network-connected printer and Sterilog software for batch documentation (via RJ45 Ethernet interface or USB flash drive).
  • Qualification
  • We offer full equipment qualification support: IQ, OQ, and DQ.
  • Touchscreen Display and Controls
  • Simple and intuitive operation. The software provides 12 program slots, 10 of which are fully customizable.
  • Customizable Parameters:
  • Program type: solids, liquids, solid waste, liquid waste.
  • Sterilization temperature: 105°C…138°C.
  • Sterilization time: 1 min…60 min.
  • Additional overheating for liquids: 0…3°C.
  • Holding time at the end of the program for liquids: 1 min…60 min.
  • Timer for programmable start time.
  • Accessories
  • Stainless steel support cabinets, as well as transport and loading carts, are available for models LabStar 100 and above.
Autoclave models and their specifications
Brochures
LITERATURE
Made on