Value; Parameter;
Model; 3950
Measurement Method; Light scattering
Flow Rate; 0.1 CFM (2.83 L/min), Accuracy: ±5% (complies with JIS B9921 and ISO21501-4)
Sampling Time; From 6 seconds to 99 minutes and 59 seconds per 1 measurement cycle
Measurement Interval; From 6 seconds to 99 minutes 59 seconds (interval between the start of the current and the start of the next repeat measurement)
Sampling Cycles; From 1 to 999 cycles or continuous measurements
Sampling Location Classification; 99 locations
Measurement Mode; Repeat, Single, Continuous, Calculation
Measurement Display Time; From 1 to 10 seconds
Measurement Display Types; Cumulative Σ, Differential Δ, and Bar graph
Maximum Measurable Concentration; 10,000,000 particles/m³ at 10% coincidence loss (complies with JIS B9921 and ISO21501-4)
Counting Efficiency; 50±20% (for 0.1 μm PSL particles), 100±10% (for 0.2 μm PSL particles) (complies with JIS B9921 and ISO21501-4)
False Count; ≤1 particle / 35 minutes
Size Resolution; ≤15% for PSL particles around 0.3 μm (complies with JIS B9921 and ISO21501-4)
Vacuum; Built-in pump
Display; 4.3-inch color LCD, resistive touch panel
Communication Interface; USB (Host: for printer and USB flash drive, Device: for PC), Ethernet, MODBUS (RS485)
Data Logging; Up to 10,000 records
Data Format; CSV
Interface Language; English, Japanese
External Power Supply; AC adapter (Input 100 to 240 V)
Operating Environment (Main Unit); 15 to 35 °C, 0 to 85% RH (non-condensing)
Dimensions; W 150 × H 163 × D 228 mm (6" W x 2.5" H x 9" D)
Weight; 3.4 kg (7.5 lbs)