User Interface and Data Management
The on-screen menu continuously guides the user through the course of the analysis. The five results are immediately displayed on the screen and can be directly printed or transferred to a LIMS. Results are stored in the instrument's internal database and can be reviewed at any time. When printing via a directly connected USB printer or a network printer, three options are available: printing data including each individual result as well as product information, date, time, user, and the instrument serial number; statistics can be appended to this information; or printing only the force-distance break curve of the last broken tablet.
Data Integrity and Audit Trail
The PTB 500 features user-friendly functions and an audit trail that logs all events in full compliance with 21 CFR Part 11 requirements. Events can be displayed directly on the instrument or printed, for instance, in PDF format for documentation purposes. Audit trail entries can be filtered by user, product, method, and event date. A detailed description is available for each event, specifying the associated user, method, and product, as well as the exact value changed, including the previous and new values.
Linear Dimension and Hardness Measurement
First, place the tablet vertically (on its edge) against the stop wall and start the thickness measurement. The measuring jaw moves forward to record the thickness of the sample. Then, the tablet is rotated to measure the width (performed if necessary). Finally, the sample is rotated once more for the combined measurement of diameter and hardness (breaking force).
Automatic or Manual Restart
Clear step-by-step instructions on the screen guide the user through each stage of the test. The analysis can be performed either in automatic restart mode, where the procedure continues independently after a configurable time interval, or in manual mode, where the user confirms the correct position of the tablet before each subsequent step.
The PTB 500 NICE instrument features a sample carousel that transfers samples between the weight and thickness measuring stations, and the combined station for measuring width, diameter (or length), and hardness. The PTB 500 NICE comes equipped with an integrated Sartorius analytical balance to measure the weight of each individual sample. For linear measurements, users can select either millimeters (mm) or inches (Inch) as the unit of measurement, while hardness can be displayed in Kiloponds (KP), Newtons (N), or Strongcobbs (Sc).
Starting a Test
The PTB 500 NICE instrument features a large, backlit color LCD. Menu navigation is performed via the touchscreen interface; it also supports the connection of a standard keyboard and mouse. A test on the PTB 500 NICE can be initiated in two ways: by starting a "Quick Test" immediately after powering on the instrument without prior system authorization, or by logging into a user account to perform an analysis based on a preconfigured testing method.
Material Transport and Testing Sequence
The samples are automatically transported to the measuring stations via a rotating carousel. First, the integrated Sartorius analytical balance measures the weight. Next, the thickness is measured, followed by the third step at the combined station for diameter (or length) and hardness testing. It is also possible to measure the sample width. In this case, the operator is prompted to position the sample accordingly on the measuring station. During diameter (or length) and hardness measurement, the jaw moves forward, touches the tablet at its highest point to measure the diameter, and immediately determines its hardness (the force required to fracture the tablet). Finally, the broken sample is discharged into the waste container.
Test Results and Reporting
When printing via a directly connected USB printer or a network printer, three options are available: printing data including each individual result as well as product information, date, time, user, and the instrument serial number; statistics can be appended to this information; or printing only the force-distance break curve of the last broken tablet. Individual measuring stations can be activated or deactivated at any time. Consequently, it is possible to test only the tablet hardness without measuring other parameters. To do this, deactivate the other parameters (e.g., thickness) by leaving the nominal value and sample quantity fields blank.
Operating Principle
A higher measuring jaw speed reduces reproducibility and often leads to overestimated results. To ensure the option of selecting an operating mode that yields results consistent with your existing equipment, choose the force mode: either linear force increase or linear speed increase, and set the same or a comparable speed. The contact force and detection force can also be modified to accommodate the specific structural characteristics of the sample. Upon contact with the sample, the instrument switches to the selected mode and linear rate of increase. Additionally, the instrument offers a capsule testing feature via a specialized procedure that allows the device to merely touch and compress the sample without fracturing it.
Calibration and Validation
The current USP pharmacopeia requires that the force sensor of a tablet hardness testing instrument be periodically calibrated across the entire measuring range (or the range used for sample analysis) with an accuracy of 1 N. All Pharma Test tablet hardness testing instruments can undergo static calibration across the entire measuring range using various certified counterweights. All instruments support verification at a minimum of three different points during calibration to confirm the linearity of the force sensor.
Technical Features
- Pharmacopeia Compliance: Tablet hardness testing in full compliance with current USP <1217> and EP <2.9.8> requirements.
- Comprehensive Testing: 5 results for the exact same sample: thickness, width, diameter (or length), hardness, and weight.
- Dual Force Mode: Equipped with both linear speed increase and linear force increase modes.
- Statistical Analysis: Automated statistical calculation of mean value, absolute and relative standard deviation, as well as minimum and maximum results.
- Connectivity: Full network capabilities and seamless direct connection to a LIMS or other network resources.
- Validation Tools: Built-in multi-point validation procedure and load cell adjustment.
- Data Visualization: Programmable printing of the force-distance break curve.
- User Interface: Intuitive graphical user interface with clear step-by-step instructions for the operator.
Advantages
- Dual Mode Flexibility: Choice between linear force increase or linear speed increase (dual-mode selection).
- Precise Control: Infinitely adjustable speed or rate of force increase.
- Data & Compliance Management: Sophisticated user and method management system, along with a results database and audit trail.
- Data Integrity: Secure password protection with customizable password complexity rules.
- Traceability: On-screen display and hard-copy printout of the time and date.
- Operational Efficiency: Quick-start feature (Test Start) to initiate analysis with minimal preparation.
- Flexible Data Export: Serial data transmission via the RS-232 interface or network integration.
- Maintenance Programs: Dedicated validation and calibration program for all measuring stations.
- Sensor Adjustment: Two-point load cell adjustment for the hardness testing station.
- Extended Validation: Multi-point validation (calibration) utilizing counterweights up to 30 or 50 kg.
- Special Applications: Program for testing soft gelatin capsules by configuring a specific target deformation distance.