Print-Out Results
The result is immediately displayed on the screen and output to a connected laser (PCL5-supported) or dot-matrix printer. Repeat this procedure until the entire series has been tested to obtain a comprehensive report. The report includes each individual result, date, time, instrument serial number, batch number of the product under test, the mean value, and the deviations for the specific test series.
Operating Principle
The driven test jaw is moved by a stepper motor toward the neck of the ampoule under test. The ampoule is placed onto a DIN/ISO-compliant support. As soon as the jaw contacts the ampoule, the instrument begins to apply the load in a linear speed increase mode until the sample breaks; the standard feed rate is 10 mm/min. At the moment of the ampoule fracture, the maximum breaking force is logged on the display and sent for printing. The test jaw then returns to its initial position, ready for the next test. Since the operator can configure an automatic restart time, the next test will initiate immediately after a new sample is placed onto the ampoule support.
The ampoule support is mounted on a movable drawer, and a Plexiglas (acrylic glass) enclosure protects the operator. An integrated lamp illuminates the work area to facilitate the correct positioning of the ampoule on the support. Interchangeable supports enable the testing of any ampoule type with volumes ranging from 1 to 30 ml.
Features
- Dual Force Application Modes: Equipped with both linear speed increase and linear force increase modes.
- Validation Support: Integrated multi-point validation procedure.
- Data Visualization: Programmable printing of the force-distance break curve.
Advantages
- Standard Compliance: Empty ampoule break strength testing in full compliance with the DIN EN ISO <9187-1/2:2010> standard.
- Precise Loading: Standard force setting featured as a linear speed increase of 10 mm/min (adjustable).
- Operational Efficiency: Automatic restart function to accelerate the testing sequence.
- Traceability: Manual data entry for time, date, and a 12-digit batch number.
- Flexible Documentation: Hard-copy documentation of all results using an external laser (PCL5) or dot-matrix printer.
- Analytical Graphics: Programmable printing of the force-distance break curve.
- Data Export: Serial data transmission via the integrated RS-232 interface.
- Maintenance Programs: Dedicated validation and calibration program for the measuring station.
- Sensor Adjustment: Two-point load cell adjustment for the strength measuring station.
- Metrological Control: Full multi-point validation (calibration) capabilities.