Full Regulatory Compliance
One of the testing criteria for the mechanical strength of tablets and cores according to USP <1216>, EP <2.9.7>, JP <14>, and other pharmacopeias is the determination of friability and abrasion. During coating, transportation, and packaging processes, tablets lose a certain portion of their weight. To measure this weight loss, the samples are counted and weighed. A friability test is then performed in accordance with the individual monographs of the respective pharmacopeia. With each revolution of the friability tester's drum, the tablets are rolled by an internal curved scoop that extends from the center of the drum to its outer wall.
Automated Discharge
Prior to starting the test, de-dust the samples. Weigh the samples and place them into any of the instrument's drums. The drums open easily for loading and cleaning. Alternatively, the discharge aperture can be used to load the samples without opening the drum housing itself. Select the testing method and start the process. A color-changing backlit LCD displays the status and analysis results.
Upon test completion, the rotation of the drums stops, and the samples are discharged into individual stainless steel collectors featuring a perforated base. These collectors are mounted on a sample collection tray, also constructed from stainless steel. The drum support shaft is completely removable. The samples must then be inspected for fractured tablets, de-dusting must be performed, and the samples reweighed. The weight difference before and after testing is defined as friability; this indicator typically should not exceed 1%.
Optional Abrasion Drums
While standard friability drums (also known as "Roche" drums) are included in the standard scope of delivery, abrasion drums are available as an option. The standard friability drum causes the samples to roll and drop during the test, whereas the abrasion drum subjects the samples to continuous mechanical stress via contact with arrow-shaped blades.
Variable Speed Settings
A silent DC gear motor rotates the drums at a fixed speed of 25 rpm or with adjustable speed options (the setting can be locked). The drum itself is constructed from Plexiglas (Perspex) and consists of two parts: the drum body and a removable cover, which opens to load and clean the drum interior when necessary. The instrument housing is manufactured from high-quality stainless steel, fully compliant with current GLP (Good Laboratory Practice) requirements. A friability drum featuring an additional anti-static coating is available as an option.
Click-Wheel Navigation and User Interface
The PTF 600 instrument is operated by using a click wheel to navigate menus on the LCD screen and by functional keys to shortcut frequent operations. The display is backlit in changing colors that show the instrument status in a familiar traffic light system (green, yellow, red).
Navigating the menu is quick and simple. The color-changing display makes it easy for the user to see when his input is required to proceed. The instrument has a method and a user management system. Methods can be programmed, edited and stored directly on the instrument. Users can be created and assigned different user levels according to their roles (administrator, operator, calibration technician).
Access to the instrument is protected by a login with username and password. Optionally a quick test menu can be accessed without logging in to the instrument first. This way tests can be performed with the instrument in case user access control and testing according to defined methods is not required.
Integrated Report Printer
The PTF 600 features an integrated report printer to document the test and calibration results. An integrated printer saves valuable bench space in your laboratory. The detailed result report includes the automatic weight loss calculation.
Data Storage and Export Options
A USB flash drive can be connected to the instrument to save and load testing methods in .csv format. Printed reports can be stored as text files. Furthermore, a USB flash drive can be utilized to install instrument firmware updates and restore factory settings without the need for a PC or any programming tool.
Technical Features
- Pharmacopeia Compliance: Fully USP <1216>, EP <2.9.7> and JP <14> compliant.
- Application: Determine friability and abrasion (optional) of uncoated tablets.
- Flexibility: Programmable for number of drum revolutions or rotating time.
- Metrological Control: Connect a balance to determine the sample weight (optional, weight can be entered manually).
- Data Transfer: Connect USB flash drive to copy methods and store test reports.
- Documentation: Integrated protocol printer.
- Access Control: User and method management system.
Advantages
- Ergonomic Design: Drum is easy to open for loading and cleaning, alternatively use discharge opening for loading.
- Clear Interface: Graphical LCD screen showing actual and target settings.
- Compact Footprint: Integrated protocol printer reduces bench space requirements.
- Comprehensive Reporting: Detailed result report including weight loss calculation and evaluation of the results.
- Operational Efficiency: Automated sample discharge at the end of a test run.
- Dual Operation Mode: Choice between number of revolutions or rotation time.
- Precise Control: Variable drum rotation speed or locked speed mode.
- Automation: Determine the sample weight by balance to automatically calculate weight loss.
- Compliance Management: Programmable qualification interval timer reminds the user to perform instrument calibration.
- Validation Support: Dedicated calibration programs and reports.
- Method Cloning: Copy methods between instruments by using a common USB flash drive.
- Data Export: Export test reports as text files.
- Standard Documentation: IQ/OQ documents included free of charge.