Full Regulatory Compliance
One of the criteria for testing the mechanical strength of tablets and cores according to USP <1216>, EP <2.9.7>, JP <14>, and other pharmacopeias is the determination of friability and abrasion. During coating, transportation, and packaging, tablets lose a certain portion of their weight. To measure this weight loss, the samples are counted and weighed. The friability test is then performed in accordance with the individual monographs of the respective pharmacopeia. The tablets tumble at each turn of the friability tester drum by means of a curved scoop that extends from the middle of the drum to its outer wall.
Automatic Discharge
Dedust the samples before starting the test. Weigh the samples and place them into any of the instrument's drums. The drums open easily for loading and cleaning. Alternatively, the discharge opening can be used to load the samples without opening the drum itself. Select the test method and start the run. The color-changing backlit LCD displays the test status and results.
Once the test is complete, the rotation of the drums stops, and the samples are discharged into individual stainless steel collectors with a perforated base. The collectors are placed on a sample collection tray, which is also made of stainless steel. The drum support axle is removable. Next, the samples must be inspected for broken tablets, dedusted, and reweighed. The difference in weight before and after the test is determined as friability; this value should typically not exceed 1%.
Optional Abrasion Drums
Friability drums (also known as "Roche" drums) are included in the standard scope of delivery, while abrasion drums are available as an option. A standard friability drum causes the samples to tumble and drop during the test, whereas an abrasion drum subjects the samples to continuous mechanical stress through contact with lamellae.
Variable Speed Settings
A silent DC geared motor drives the drums at a fixed speed of 25 rpm or at a selectable speed (the setting can be locked). The drum itself is made of Perspex and consists of two parts: the drum body and a removable cover that opens for loading and cleaning the inside of the drum when required. The instrument itself is manufactured from high-grade stainless steel in full compliance with current GLP (Good Laboratory Practice) requirements. A friability drum with an additional anti-static coating is available as an option.
User Interface with Rotary Knob Navigation
The PTF 600 is operated via a rotary knob (scroll wheel) for navigating the LCD menu and function keys for quick access to frequently performed operations. The display features a color-changing backlight that indicates the instrument's status using a familiar traffic light principle (green, yellow, red).
Menu navigation is fast and straightforward. The color-changing display backlight allows the user to easily identify when operator action is required to proceed. The instrument is equipped with a method and user management system. Methods can be programmed, edited, and stored directly in the instrument's memory. User accounts can be created with different access levels assigned according to their roles (Administrator, Operator, Calibration Specialist).
Access to the instrument is protected by a login procedure requiring a username and password. Optionally, a quick test menu can be accessed without prior system login. This enables testing to be carried out in cases where user access control and strictly defined test methods are not required.
Built-in Report Printer
The PTF 600 is equipped with a built-in report printer for documenting test and calibration results. The integrated printer saves valuable bench space in the laboratory. The detailed results report includes automatic weight loss calculation.
Data Storage Capabilities
A USB flash drive can be connected to the instrument to save and load methods in .csv format. Report printouts can be saved as text files. Additionally, the flash drive can be used to install instrument firmware updates and restore factory defaults without the need for a PC or external programmer.
Features- Full compliance with USP <1216>, EP <2.9.7>, and JP <14> pharmacopeia requirements
- Friability and abrasion (optional) testing of uncoated tablets
- Programmable by number of drum revolutions or rotation time
- Balance connection for sample weight determination (optional; weight can be entered manually)
- USB flash drive connection for copying methods and storing test reports
- Built-in protocol printer
- User and method management system
Advantages- Drum opens easily for loading and cleaning; the discharge opening can also be used for loading
- Graphical LCD screen displaying current and target settings
- Built-in protocol printer saves laboratory bench space
- Detailed results report, including weight loss calculation and result evaluation
- Automatic sample discharge upon test completion
- Two operating modes: by number of revolutions or by rotation time
- Adjustable drum rotation speed or locked speed mode
- Sample weight determination via a balance for automatic weight loss calculation
- Programmable qualification interval timer reminds the user when instrument calibration is due
- Dedicated calibration programs and reports
- Copying methods between instruments using a standard USB flash drive
- Export of test reports as text files
- IQ/OQ documentation included free of charge