The instrument fully complies with the current European Pharmacopoeia (EP) and German Pharmacopoeia (DAB) requirements. In accordance with the pharmacopeial monograph, three samples are tested simultaneously. The total test duration can be preset within a range of 1 minute to 10 hours. During suppository testing, the sample basket assemblies automatically rotate 180 degrees every 10 minutes, as strictly required by the relevant pharmacopeial monograph.
Test Procedure
The three-sample testing basket assembly is placed into a 12-liter Plexiglas water bath. Preset the total test duration and the basket rotation cycle (typically 10 minutes), then start the test. Throughout the entire analysis runtime, the basket assembly will automatically rotate 180 degrees at the specified intervals. Generally, a suppository completely disintegrates within 30 minutes. When testing pessaries, no rotation cycle time is entered; instead, the samples are placed onto the upper part of the basket assembly, which is then covered with the included glass plate.
The integrated thermostatically controlled heating system heats the water in the bath to 37 °C or another required target temperature. System heating and water circulation are driven by a 1000 W heater and a high-capacity pump. The heating system design is specifically engineered to ensure absolute protection against the intrusion of fats or active pharmaceutical ingredients (APIs). For cleaning purposes, the water bath, sample holder, and testing basket assemblies can be completely removed from the instrument.