Tablet disintegration testers are widely utilized throughout the pharmaceutical industry to evaluate the disintegration characteristics of finished dosage forms and ensure the quality control of various solid oral medications. All tablet and capsule disintegration testers manufactured by Pharma Test fully comply with the current United States Pharmacopeia (USP) and European Pharmacopoeia (EP) specifications. They support the use of both Type "A" and Type "B" basket assemblies for the analysis of standard or larger-sized samples. The instrument housing is constructed from stainless steel, ensuring stricter compliance with GMP requirements compared to painted metal structures. Since these instruments are frequently operated in Quality Control (QC) and In-Process Control (IPC) laboratories where most other equipment is also made of stainless steel, Pharma Test instruments seamlessly integrate into the workflow environment.
The disintegration basket assembly, which holds the glass tubes and discs, is inserted into a metal holder and secured via a magnetic lock. The instrument automatically detects whether the basket is properly installed. A backlit liquid crystal display (LCD) provides operational status updates, such as the elapsed testing time. The disintegration basket assembly performs reciprocal vertical movements (up and down) with a stroke height of 55 mm (in each direction) at a frequency of 30 cycles per minute. At the start of the test, it smoothly lowers into the dissolution medium.
The instrument's heating system, equipped with a circulation pump, heats the water within the Plexiglas water bath to the configured target temperature (e.g., 37 °C). A secondary sensor acts as a safety device, shutting off the heater if the temperature exceeds the permissible limit. A flow sensor, thermal fuse, and thermo-switch serve as additional safety elements to protect the system from overheating. The target bath temperature can be freely adjusted. The temperature of the testing medium can be verified at any time using the included PT100 temperature probe.
Depending on the Standard Operating Procedure (SOP), the test is terminated either automatically upon reaching the preset time runtime, or manually by the operator once visual confirmation is made that all samples have completely disintegrated. Upon completion, the basket assembly stops at its uppermost mechanical position. The standard Type "A" basket assembly comes complete with six glass tubes and six plastic discs.
Optional Type "B" Basket Assemblies
The three-tube Type "B" basket assembly, compliant with European Pharmacopoeia (EP <2.9.1.2>) and United States Pharmacopeia (USP <2040>) specifications, can be utilized for the analysis of larger-sized samples. It is supplied ready-to-use, complete with three glass tubes and three plastic discs, and can be easily interchanged.
Additionally, other specialized basket assemblies are available for testing capsules or vaginal tablets.
Click-Wheel Navigation and User Interface
The PTZ 100 and PTZ 300 instruments are operated via a click-wheel for menu navigation on the LCD, along with functional keys for quick access to frequently performed operations. The display features a variable-color backlight that indicates the instrument status using a traffic-light system (green, yellow, red). Menu navigation is fast and simple. The color-changing display allows the operator to easily identify when manual intervention is required to proceed with the workflow.
Method and User Management System
The instruments feature an integrated method and user management system. Testing methods can be programmed, edited, and stored directly within the instrument's internal memory. User accounts can be created within the system, with various access levels assigned according to user roles (such as method developer or standard user, with only one administrator account permitted).
Access to the instrument is secured via a login procedure requiring a username and password. Additionally, a "Quick Test" menu is available, which can be accessed without prior system authorization. This allows operators to perform tests when user access control and adherence to strictly defined methods are not required.
Integrated Report Printer
The PTZ 100 and PTZ 300 instruments feature an integrated report printer to document test and calibration results. An integrated printer saves valuable bench space in your laboratory.
Cleaning and Maintenance
For convenient cleaning and maintenance, the Plexiglas water bath can be easily removed. Quick-disconnect couplings connect the tubing to the heating system and automatically seal on both sides upon disconnection.
Data Storage and Export
A USB flash drive can be connected to the instrument to save and load testing methods in .csv format. Printed reports can be saved as text files. Test results can be transferred to external systems, such as a LIMS (Laboratory Information Management System), via the serial interface. Furthermore, a USB flash drive can be utilized to install instrument firmware updates and restore factory settings without the need for a PC or any external programming tools.
PT-Node Network Adapter for Printing and Data Transfer
The PT-Node is an adapter that enables the simultaneous connection of up to two Pharma Test instruments to a network via a wired LAN connection. This allows you to print test results from the instrument through a web browser on any local or network printer. Additionally, test results can be transferred from the instruments to external systems located within the same network. The PT-Node adapter supports operation with both the PTZ 100 and PTZ 300 instruments.
Calibration
The PTZ 100 and PTZ 300 instruments feature dedicated calibration programs for temperature, basket stroke, and timer functions. Calibration results can be documented via a hard-copy printout. The system guides the user step-by-step through the entire calibration process, after which a report is automatically printed.
Reports are stored within the instrument's internal memory and can be printed at a later time. A programmable qualification interval timer reminds the user to perform instrument calibration when the scheduled due date arrives. Upon each power-up, the instrument performs a quick self-test to verify operational readiness. The self-test result can also be documented using the built-in printer (available upon request).
Technical Specifications
- Pharmacopeia Compliance: Full compliance with current USP <701/2040>, EP <2.9.1 / Test A and B>, and JP requirements.
- Configurations: Available in single-station or triple-station configurations (PTZ 100 and PTZ 300 models).
- Application: Disintegration time determination for tablets, capsules, and other dosage forms.
- Access Control: Integrated user management system.
- Method Management: Integrated testing method management system.
- Temperature Control: External temperature probe for monitoring the testing medium temperature.
- Calibration Tools: Built-in calibration programs.
- Data Transfer: USB flash drive connectivity for copying methods and saving test reports.
- Documentation: Built-in printer for protocol and report generation.
- Maintenance Features: Self-sealing quick-disconnect couplings for water bath connection.
Advantages
- Clear Interface: Graphical LCD displaying actual and set parameters.
- Compact Design: Built-in protocol printer saves valuable laboratory bench space.
- Comprehensive Reporting: Result report includes disintegration times, actual temperature, and the number of non-disintegrated samples.
- Data Integrity: Password-protected access via the user management system.
- Compliance Management: Programmable qualification interval timer reminding the user of upcoming instrument calibration deadlines.
- Validation Support: Specialized calibration programs and report templates.
- Efficiency: Method cloning/copying between instruments via a standard USB flash drive.
- Data Export: Export of test reports as text files.
- LIMS Integration: Seamless transfer of test results to external laboratory information systems (such as LIMS).
- Standard Documentation: Installation and Operational Qualification (IQ/OQ) documentation is included in the standard scope of delivery (provided free of charge).