Universal Oven UF30plus

The U series universal laboratory drying oven is the classic Memmert equipment for thermal processing in science, research, and industrial material testing. A technologically advanced masterpiece made of high-quality, hygienic, and easy-to-clean stainless steel, it fully meets all requirements regarding ventilation, control technology, overtemperature protection, and precisely tuned heating systems.


Safety
  • Temperature Control (Protection): Mechanical temperature limiter TB, protection class 1 according to DIN 12880, to switch off the heating at approx. 20 °C above nominal temperature.
  • Temperature Control (Choice): Overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display.
  • AutoSAFETY Function: Additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range. Alarm in case of over- or undertemperature, heating is switched off in case of overtemperature.
  • Self-Diagnostic System: For fault analysis.
  • Alarm System: Visual and acoustic.
Standard Equipment
  • Works Calibration Certificate: Calibration at +160 °C.
  • Door: Fully insulated stainless steel door with 2-point locking (compression door lock).
  • Interior Elements: 1 electropolished stainless steel grid.
Chamber Dimensions and Specifications
  • Chamber Dimensions (W x H x D): 400 x 320 x 250 mm (depth reduced by 39 mm for fan).
  • Inner Chamber: Easy-to-clean stainless steel inner chamber, reinforced by deep drawn ribbing, with integrated and protected large-area heating on all four sides.
  • Volume: 32 L
  • Max. Number of Internals/Grids: 3 pcs.
  • Max. Chamber Load: 60 kg
  • Max. Load per Grid: 20 kg
Housing and Installation
  • Housing: Textured stainless steel.
  • Housing Dimensions (W x H x D): 585 x 704 x 434 mm (depth + 56 mm for door handle).
  • Housing Rear Wall: Zinc-plated steel.
Electrical Parameters
  • Option 1 (Voltage / Power): 230 V, 50/60 Hz / approx. 1600 W
  • Option 2 (Voltage / Power): 115 V, 50/60 Hz / approx. 1600 W
Ambient and Installation Conditions
  • Installation: The distance between the wall and the rear wall of the equipment must be at least 15 cm. The clearance to the ceiling must be at least 20 cm, and the side clearance to walls or adjacent equipment must be at least 5 cm.
  • Installation Altitude: Max. 2,000 m above sea level.
  • Ambient Temperature: +5 °C to +40 °C.
  • Ambient Humidity: Max. 80% relative humidity, non-condensing.
  • Overvoltage Category: II.
  • Pollution Degree: 2.
Packaging and Shipping Data
  • Shipping Information: The equipment must be transported in an upright position.
  • Customs Tariff Code (HS Code): 8419 89 98
  • Country of Origin: Federal Republic of Germany.
  • WEEE Registration Number: DE 66812464
  • Packaging Dimensions (W x H x D): approx. 660 x 890 x 650 mm (including cardboard box).
  • Net Weight: approx. 45 kg
  • Gross Weight: approx. 61 kg
TECHNICAL SPECIFICATIONS
Temperature
  • Operating Temperature Range: At least 5 °C (for UN/UNplus/UNm/UNmplus models) or 10 °C (for UF/UFplus/UFm/UFmplus models) above ambient temperature up to +300 °C.
  • Temperature Setting Accuracy: Up to 99.9 °C: 0.1 °C / from 100 °C: 0.5 °C.
  • Temperature Setting Range: +20 °C to +300 °C.
  • Temperature Sensors: 2 Pt100 sensors, DIN class A in 4-wire circuit (for mutual monitoring and redundancy in case of a fault).
Control Technology
  • Control System: ControlCOCKPIT with two displays (TwinDISPLAY). Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition color TFT displays.
  • Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days.
  • SetpointWAIT Function: Process time does not start until the set temperature is reached.
  • Calibration: Three freely selectable temperature values.
  • Adjustable Parameters: Temperature (Celsius or Fahrenheit), fan speed, air flap position, program time, time zones, daylight saving time.
Ventilation
  • Fan: Forced air circulation by quiet air turbine, adjustable in 10 % steps for each segment individually.
  • Fresh Air: Admixture of pre-heated fresh air by electronically adjustable air flap.
  • Vent: Vent connection with restrictor flap.
Communication
  • Documentation: Program stored in case of power failure.
  • Programming: AtmoCONTROL software on a USB stick for programming, managing, and transferring programs via Ethernet interface or USB port.



Product specification
LITERATURE
Made on
Tilda