Laboratory oven UN750

The universal laboratory oven U is Memmert’s classic appliance for temperature control in science, research and material tests in industry.
The technologically perfected masterpiece made of high-quality, hygienic, easy-to-clean stainless steel.
Leaves nothing to be desired in terms of ventilation and control technology, overtemperature protection and precisely tuned heating technology.

Safety
Temperature control (protection): Adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880, to switch off heating at approx. 20 °C above nominal temperature.
Autodiagnostic system: For fault analysis.
Standard equipment
Works calibration certificate: Calibration at +160 °C.
Doors: Fully insulated stainless steel doors with 2-point locking (compression door lock).
Internals: 2 electropolished stainless steel grids.
Chamber dimensions and specifications
Interior: Easy-to-clean interior made of stainless steel, reinforced by deep-drawn ribbing, with integrated and protected large-area heating on four sides.
Volume: 749 l.
Chamber dimensions (W x H x D): 1040 x 1200 x 600 mm.
Max. number of internals/grids: 14 pcs.
Max. loading of chamber: 300 kg.
Max. loading per internal/grid: 30 kg.
Housing and installation
Housing: Textured stainless steel.
Housing dimensions (W x H x D): 1224 x 1720 x 784 mm (depth + 56 mm for door handle).
Installation: On lockable castors.
Rear panel of housing: Zinc-plated steel.
Electrical data
Voltage / Power: 400 V and 3x 230 V without neutral, 50/60 Hz / approx. 7000 W.
Ambient and installation conditions
Installation: The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm, and the side clearance from walls or nearby appliances must not be less than 5 cm.
Installation altitude: Max. 2,000 m above sea level.
Ambient temperature: +5 °C to +40 °C.
Humidity: Max. 80 % rh, non-condensing.
Overvoltage category: II.
Pollution degree: 2.
Packing and shipping data
Transport information: The appliance must be transported upright.
Customs tariff number: 8419 8998
Country of origin: Federal Republic of Germany.
WEEE-Reg.-No.: DE 66812464.
Packaging dimensions (W x H x D): approx. 1330 x 1910 x 1050 mm (incl. carton).
Net weight: approx. 217 kg.
  • Gross weight: approx. 288 kg.
TECHNICAL SPECIFICATIONS
Temperature
Working temperature range: At least 5 °C (for UN/UNplus/UNm/UNmplus models) or 10 °C (for UF/UFplus/UFm/UFmplus models) above ambient temperature up to +300 °C.
Setting accuracy temperature: Up to 99.9 °C: 0.1 °C / from 100 °C: 0.5 °C.
Setting temperature range: +20 °C to +300 °C.
Temperature sensor: 1 Pt100 sensor DIN Class A in 4-wire-circuit.
Control technology
ControlCOCKPIT: SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display.
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days.
Function SetpointWAIT: The process time does not start until the set temperature is reached.
Calibration: Three freely selectable temperature values.
Adjustable parameters: Temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime.
Ventilation
Convection: Natural convection.
Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap.
Vent: Vent connection with restrictor flap.
Communication
Documentation: Programme stored in case of power failure.
Programming: AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).