Safety control systemMechanical temperature limiter (TB): Protection class 1 according to DIN 12880, switches off heating at approx. 10 °C above nominal temperature.
AutoSAFETY system: Additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value within a preset tolerance range; alarm in case of over- or undertemperature, heating switched off in case of overtemperature, compressor switched off in case of undertemperature.
Autodiagnostic system: Integrated fault diagnosis for temperature control.
Alarm: Visual and acoustic.
Heating conceptDirect heating VO: Fuzzy-supported MLC microprocessor controller (Multi-Level-Controlling) adapts its performance to the volume (local temperature sensing) for each thermoshelf.
Shelf sockets: 2 thermoshelf sockets on the rear wall (on 1st and 3rd level).
Standard equipmentCalibration certificate: Works calibration certificate for +160 °C at 20 mbar for each supplied thermoshelf.
Door: Full-sight glass door; inside spring-mounted 15 mm safety glass pane, outside pane with anti-splinter screen.
Internals: 1 thermoshelf made of aluminium, material 3.3547 (ASTM B209) with integrated large-area heating.
Interior material: Hermetically welded interior made of particularly corrosion-resistant stainless steel, material 1.4404.
Chamber construction: Removable mountings on the sides made of stainless steel (material 1.4404) with guides for thermoshelves; top diffuser to prevent turbulence when venting.
Dimensions and loadingVolume: 101 l.
Internal dimensions (W x H x D): 545 x 465 x 400 mm.
Max. number of internals: 2 pcs.
Max. total loading of chamber: 60 kg.
Max. loading per internal: 20 kg.
Housing: Textured stainless steel.
External dimensions (W x H x D): 710 x 767 x 550 mm (depth + 38 mm for door handle).
Rear panel of housing: Zinc-plated steel.
Electrical dataVoltage and electrical load: At maximum equipment from 230 V, 50/60 Hz
(exact wattage value is omitted in the source text).
Installation and ambient conditionsAmbient temperature: +5 °C to +40 °C.
Installation requirements: The distance between the wall and the rear of the appliance must be at least 15 cm; the clearance from the ceiling must not be less than 20 cm; the side clearance from walls must not be less than 8 cm.
Humidity: Relative humidity max. 80 %, non-condensing.
Overvoltage category: II.
Pollution degree: 2.
Packing and shipping dataTransport information: The appliance must be transported upright.
Customs tariff number: 8419 8998.
Country of origin: Federal Republic of Germany.
WEEE-Reg.-No.: DE 66812464.
Packaging dimensions (W x H x D): approx. 820 x 970 x 670 mm.
Net weight: approx. 110 kg.
- Gross weight (in carton): approx. 135 kg.