General specification; 2-glove; 3-glove; 4-glove; 5-glove; Transfer chamber;
Nominal process chamber size (width); 1200 mm; 1600 mm; 2000 mm; 2400 mm; 680 mm
External dimensions (WxDxH) with 1 transfer chamber; 1990 x 924 x 2275 mm; 2390 x 924 x 2275 mm; 2790 x 924 x 2275 mm; 3190 x 924 x 2275 mm; N/A
Internal chamber dimensions (WxDxH); 1200 x 633 x 842 mm; 1600 x 633 x 842 mm; 2000 x 633 x 842 mm; 2400 x 633 x 842 mm; 682 x 532 x 842 mm
Chamber pressure; Negative pressure; Negative pressure; Negative pressure; Negative pressure; Negative pressure
Airflow type; Turbulent flow; Turbulent flow; Turbulent flow; Turbulent flow; Turbulent flow
Filter efficiency (Inlet); Push-push 99.995% @ 0.3µm MPPS; Push-push 99.995% @ 0.3µm MPPS; Push-push 99.995% @ 0.3µm MPPS; Push-push 99.995% @ 0.3µm MPPS; Push-push 99.995% @ 0.3µm MPPS
Filter efficiency (Outlet); Push-push 99.995% @ 0.3µm MPPS; Push-push 99.995% @ 0.3µm MPPS; Push-push 99.995% @ 0.3µm MPPS; Push-push 99.995% @ 0.3µm MPPS; Cartridge 99.98% @ 0.3µm MPPS
Exhaust volume (Normal operation); 20 m³/h @ 300 Pa; 30 m³/h @ 300 Pa; 40 m³/h @ 300 Pa; 50 m³/h @ 300 Pa; N/A
Exhaust volume (Emergency/breach); 100 m³/h @ 900 Pa; 100 m³/h @ 900 Pa; 100 m³/h @ 900 Pa; 100 m³/h @ 900 Pa; N/A
Illuminance level; ≥ 500 Lux; ≥ 500 Lux; ≥ 500 Lux; ≥ 500 Lux; N/A
Noise level; ≤ 75 dBA; ≤ 75 dBA; ≤ 75 dBA; ≤ 75 dBA; ≤ 75 dBA
Material: Inner chamber; SS316L, 4.0 mm, Ra ≤ 0.40 µm; SS316L, 4.0 mm, Ra ≤ 0.40 µm; SS316L, 4.0 mm, Ra ≤ 0.40 µm; SS316L, 4.0 mm, Ra ≤ 0.40 µm; SS316L, 4.0 mm, Ra ≤ 0.40 µm
Material: Outer chamber; SS304, 4.0 mm, Ra ≤ 0.60 µm; SS304, 4.0 mm, Ra ≤ 0.60 µm; SS304, 4.0 mm, Ra ≤ 0.60 µm; SS304, 4.0 mm, Ra ≤ 0.60 µm; SS304, 4.0 mm, Ra ≤ 0.60 µm
Door seals; Silicone, inflatable, 21 CFR 177.2600; Silicone, inflatable, 21 CFR 177.2600; Silicone, inflatable, 21 CFR 177.2600; Silicone, inflatable, 21 CFR 177.2600; Silicone, inflatable, 21 CFR 177.2600
Electrical requirements; 110-120V / 220-240V (1-phase) / 480V (3-phase), 50/60 Hz; 110-120V / 220-240V (1-phase) / 480V (3-phase), 50/60 Hz; 110-120V / 220-240V (1-phase) / 480V (3-phase), 50/60 Hz; 110-120V / 220-240V (1-phase) / 480V (3-phase), 50/60 Hz; 110-120V / 220-240V (1-phase) / 480V (3-phase), 50/60 Hz
Compressed air requirements; Min. 6 bar, max. 12 bar, ≥ 50 L/min; Min. 6 bar, max. 12 bar, ≥ 50 L/min; Min. 6 bar, max. 12 bar, ≥ 50 L/min; Min. 6 bar, max. 12 bar, ≥ 50 L/min; Min. 6 bar, max. 12 bar, ≥ 50 L/min
Chamber cleanliness; ISO Class 7; ISO Class 7; ISO Class 7; ISO Class 7; ISO Class 7
Air exchange rate; > 30 ACH; > 30 ACH; > 30 ACH; > 30 ACH; > 30 ACH
Control system; 12" HMI with PLC (option: SCADA); 12" HMI with PLC (option: SCADA); 12" HMI with PLC (option: SCADA); 12" HMI with PLC (option: SCADA); 12" HMI with PLC (option: SCADA)