Ideally suited for temperature control of ointments, emulsions, samples, plates, and nutrient media in laboratory environments, as well as for warm-keeping and calibration in the industrial sector.
Heating System
Heating Principle: Large-area, corrosion-resistant heating element on the outside of the inner chamber.
Standard Equipment
Cover: Gabled stainless steel cover.
Drainage System: Innovative drainage system — easy emptying.
Timer: Digital timer from 1 min to 99 hours 59 min, with visual and acoustic alarm.
Chamber Dimensions and Specifications
Inner Chamber: Deep-drawn stainless steel tank (steel grade W.St. 1.4301) without corners or edges for easy cleaning, including a central drain (no obstructing internal components).
Chamber Dimensions (L x W x H): 472 x 267 x 200 mm
Volume: 23.1 L
Min. Liquid Level: 20 mm
Max. Liquid Level: 164 mm
Housing and Installation
Housing: Textured stainless steel.
Housing Dimensions (L x W x H): 749 x 349 x 425 mm
Installation: Ventilation openings on the left and rear sides must remain unobstructed. Clearance to the wall must be at least 100 mm on the sides and at least 150 mm at the rear. Minimum distance from the top of the bath to the ceiling is 750 mm.
Packaging Dimensions (L x W x H): approx. 849 x 449 x 525 mm (including cardboard box).
Shipping Information: The equipment must be transported in an upright position.
Customs Tariff Code (HS Code): 8419 89 98
Country of Origin: Federal Republic of Germany
WEEE Registration Number: DE 66812464
Net Weight: approx. 20 kg
Gross Weight: approx. 23 kg
TECHNICAL SPECIFICATIONS
Temperature
Temperature Sensor: Digital temperature sensor mounted on the outside of the bath.
Units of Measurement:
Setting in Celsius / Fahrenheit.
Operating Temperature Range: At least 5 °C above ambient temperature up to +100 °C (with cover, depending on atmospheric pressure).
Temperature Setting Range: +10 °C to +100 °C.
Setting Accuracy: 0.1 °C.
Control Technology
Calibration: 2-point calibration.
Controller: Intuitive 3.5-inch color graphic touchscreen display.
Indication: Digital display of setpoint and actual temperature, as well as (remaining) program time.
Safety
Temperature Control (Protection): Two-stage safety system to prevent overheating (pressure sensor for liquid level monitoring, thermal fuse to cut off heating in case of an error).
Alarm System: Individually adjustable overtemperature alarm (visual and acoustic).
Temperature Limiter: Independent temperature limiter for automatic heating cutoff in the event of a fault.
Self-Diagnostic System: Microprocessor-based PID temperature controller with an integrated self-diagnostic system and fault indicator.