Water Bath WTB50

Ideally suited for temperature control of ointments, emulsions, samples, plates, and nutrient media in laboratory environments, as well as for warm-keeping and calibration in the industrial sector.



Heating System
  • Heating Principle: Large-area, corrosion-resistant heating element on the outside of the inner chamber.
Standard Equipment
  • Cover: Gabled stainless steel cover.
  • Drainage System: Innovative drainage system — easy emptying.
  • Timer: Digital timer from 1 min to 99 hours 59 min, with visual and acoustic alarm.
Chamber Dimensions and Specifications
  • Inner Chamber: Deep-drawn stainless steel tank (steel grade W.St. 1.4301) without corners or edges for easy cleaning, including a central drain (no obstructing internal components).
  • Chamber Dimensions (L x W x H): 592 x 472 x 200 mm
  • Volume: 51 L
  • Min. Liquid Level: 20 mm
  • Max. Liquid Level: 164 mm [1]
Housing and Installation
  • Housing: Textured stainless steel.
  • Housing Dimensions (L x W x H): 868 x 554 x 425 mm
Electrical Parameters
  • Option 1 (Voltage / Power): 230 V, 50/60 Hz / approx. 2800 W
  • Option 2 (Voltage / Power): 115 V, 50/60 Hz / approx. 1840 W
Ambient and Installation Conditions
  • Installation: Ventilation openings on the left and rear sides must remain unobstructed. Clearance to the wall must be at least 100 mm on the sides and at least 150 mm at the rear. Minimum distance from the top of the bath to the ceiling is 750 mm.
  • Ambient Temperature: +5 °C to +40 °C
  • Ambient Humidity: Max. 80% relative humidity, non-condensing.
  • Overvoltage Category: II
  • Pollution Degree: 2
Packaging and Shipping Data
  • Packaging Dimensions (L x W x H): approx. 968 x 654 x 525 mm (including cardboard box).
  • Shipping Information: The equipment must be transported in an upright position.
  • Customs Tariff Code (HS Code): 8419 89 98
  • Country of Origin: Federal Republic of Germany
  • WEEE Registration Number: DE 66812464
  • Net Weight: approx. 30 kg
  • Gross Weight: approx. 37 kg
TECHNICAL SPECIFICATIONS
Temperature
  • Temperature Sensor: Digital temperature sensor mounted on the outside of the bath.
  • Units of Measurement: Setting in Celsius / Fahrenheit.
  • Operating Temperature Range: At least 5 °C above ambient temperature up to +100 °C (with cover, depending on atmospheric pressure).
  • Temperature Setting Range: +10 °C to +100 °C.
  • Setting Accuracy: 0.1 °C.
Control Technology
  • Calibration: 2-point calibration.
  • Controller: Intuitive 3.5-inch color graphic touchscreen display.
  • Indication: Digital display of setpoint and actual temperature, as well as (remaining) program time.
Safety
  • Temperature Control (Protection): Two-stage safety system to prevent overheating (pressure sensor for liquid level monitoring, thermal fuse to cut off heating in case of an error).
  • Alarm System: Individually adjustable overtemperature alarm (visual and acoustic).
  • Temperature Limiter: Independent temperature limiter for automatic heating cutoff in the event of a fault.
  • Self-Diagnostic System: Microprocessor-based PID temperature controller with an integrated self-diagnostic system and fault indicator.



Product specification